Lupus (13)
Plasma (9)
Lyme Disease (8)
HIV (6)
Outbreak (6)
SLE (6)
symptoms (6)
toxoplasmosis (6)
Hepatitis A (5)
Lupus Paid (5)
Lupus plasma (5)
Lupus Research (5)
Lupus study (5)
Lyme (5)
Celiac Research (4)
paid (4)
Celiac study (3)
gondii (3)
Lyme treatment (3)
AIDS (2)
Allergic reaction (2)
Allergies (2)
Arthritis (2)
Arthritis Research (2)
Borrelia Burgdorferi (2)
celiac disease (2)
celiac paid plasma (2)
CMV (2)
Cytomegalovirus (2)
disease (2)
donation (2)
EBV (2)
Hepatitis (2)
HIV Research (2)
lyme CDC (2)
Lyme community help (2)
Lyme controversy (2)
Lyme cure (2)
Lyme Disease infection (2)
Lyme Disease Introduction (2)
Lyme Disease Tests (2)
lyme long term Antibiotics (2)
lyme paid plasma research (2)
Lyme strains (2)
Lyme testing (2)
Lyme testing issues (2)
migrans (2)
mono (2)
mononucleosis (2)
panels (2)
Plasma Donation (2)
research (2)
Rheumatoid (2)
seroconversion (2)
Toxoplasmosis Gondii (2)
toxoplasmosis Paid (2)
toxoplasmosis symptoms (2)
ACCURUN (1)
Allergic (1)
Allergy plasma (1)
allergy research (1)
Allergy symptoms (1)
All-Natural Home Remedies (1)
ANA positive (1)
Anti Nuclear Antibody (ANA) (1)
anti phospholipid antibody (APS) (1)
antibodies (1)
antibody (1)
anticoagulant (1)
Anticoagulation (1)
APS (1)
Arthritis Paid (1)
Arthritis plasma (1)
Arthritis study (1)
assay (1)
autoimmune disorder (1)
babesia (1)
babesiosis (1)
Bioassay (1)
Borgdorferi (1)
Borrelia (1)
borreliosis (1)
bullseye rash (1)
California (1)
CAPRISA (1)
Cardiolipin (1)
cats (1)
CD4 (1)
CD8 (1)
Celiac (1)
Celiac diagnosis (1)
celiac disease plasma (1)
Celiac Paid (1)
celiac paid plasma donation (1)
Celiac plasma (1)
celiac vaccine (1)
Chlamydia trachomatis (1)
cholera (1)
Clinical (1)
CMV information (1)
CMV prevention (1)
CMV signs (1)
CMV symptoms (1)
CMV treatment (1)
coeliac (1)
Controls (1)
Cryopreserved (1)
CT/NG (1)
cure (1)
dengue (1)
diagnosis (1)
donate Allergy plasma (1)
Donate Plasma (1)
dsDNA Antibody (1)
E. coli (1)
Ebstein Barr Virus (1)
ELISPOT (1)
epidemiology (1)
FDA (1)
Flu (1)
Food Allergies (1)
Functionality (1)
Get paid (1)
gluten free diet (1)
Guidance (1)
HAVM (1)
Hep A (1)
HEPA (1)
Hepatitis A Paid (1)
Hepatitis A plasma (1)
Hepatitis A prevention (1)
Hepatitis A Research (1)
Hepatitis A study (1)
hepatitis a symptoms (1)
hepatitis a treatment (1)
Hepatitis Positive (1)
Hepatitis symptoms (1)
Hepatitis-A (1)
Hepatitus A (1)
Histone (1)
History of Lupus Erythematosus (1)
HIV Paid (1)
HIV paid plasma donation (1)
HIV plasma (1)
HIV study (1)
HIV Type (1)
HTLV (1)
HTLV-I (1)
HTLV-II (1)
immune (1)
Jo1 (1)
Kid allergies (1)
Lupus antibodies (1)
Lyme Disease Plasma (1)
Michigan (1)
Mitochondrial (1)
Mono college (1)
Mono positive (1)
Mono prevention (1)
Mono symptoms (1)
Mono treatment (1)
mononucleosis Paid (1)
mononucleosis plasma (1)
mononucleosis Research (1)
mononucleosis study (1)
Multi-Analyte (1)
Mumps (1)
Natural Remedies (1)
Neisseria gonorrhoeae (1)
Osteomalacia (1)
outbreaks (1)
Paid Plasma (1)
paid plasma allergy donation (1)
Paid Plasma Donation (1)
Paid Plasma Donation Study (1)
Paid Research (1)
paid study (1)
Pathogen (1)
PBMC (1)
Pertussis (1)
pet allergies (1)
phospholipid (1)
Plasmapheresis (1)
Rabbies (1)
Rapid (1)
Recovery (1)
Rheumatoid Arthritis (1)
Rickets (1)
SeraCare (1)
severe allergies (1)
SLE antibodies (1)
SSA (1)
SSB (1)
study (1)
Syphilis (1)
system (1)
Systemic lupus erythematosus (1)
tenofovir (1)
test sensitivity (1)
Testing (1)
tick (1)
tickborne illness (1)
Ticks in Tennessee (1)
Toxoplasma (1)
toxoplasmosis plasma (1)
toxoplasmosis plasma donation (1)
toxoplasmosis Research (1)
toxoplasmosis study (1)
Toxoplasmosis Wiki (1)
Toxoplasmosis Wikipedia (1)
transmission (1)
treatment (1)
Treponema pallidum (1)
Twitter (1)
Validating (1)
Validation (1)
Viability (1)
Vitamin D Deficiency (1)
what is heaptitis A (1)
What is Lupus (1)
white blood cell (1)
Whooping Cough (1)
Wiki (1)
Wikipedia (1)
zumba (1)
Aug 17

Written by: Emily Sherman
8/17/2010 8:30 AM  RssIcon       Share

 

If you have an autoimmune, infectious, or other disease, researchers need your help.  Donate plasma and receive $400/donation for your time.  Visit www.idonateplasma.com for details.

Source PDF:
http://www.seracare.com/Portals/0/Papers/Guidance%20for%20Industry.pdf

Disease Diagnosed?  Get $400 to donate plasma for research: www.idonateplasma.com our

 

Guidance for Industry

Revised Recommendations Regarding Invalidation of Test Results of Licensed and 510(k) Cleared Bloodborne Pathogen Assays Used to Test Donors

Additional copies are available from:

Office of Communication, Training and Manufacturers Assistance (HFM-40)

1401 Rockville Pike, Rockville, MD 20852-1448

(Tel) 1-800-835-4709 or 301-827-1800

http://www.fda.gov/cber/guidelines.htm

U.S. Department of Health and Human Services Food and Drug Administration

Center for Biologics Evaluation and Research (CBER) July 2001

 

GUIDANCE FOR INDUSTRY

Revised Recommendations Regarding Invalidation of Test Results of Licensed and 510(k) Cleared Bloodborne Pathogen Assays Used to Test Donors

This guidance document represents FDA’s current thinking on invalidating test results based on the Clinical Laboratory Improvement Act of 1988 (CLIA) required control reagents. It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statutes and regulations.

 

I. INTRODUCTION

To reduce infectious disease transmission by blood and blood products, donor samples from blood donations are tested for markers of pathogenic bloodborne infections, including antibodies, antigens, and nucleic acids that may indicate the presence of etiologic agents such as human immunodeficiency virus (HIV), hepatitis B virus (HBV), hepatitis C virus (HCV), human T-cell lymphotropic virus (HTLV), cytomegalovirus (CMV), syphilis, and others. The validity of screening and supplemental (confirmatory) test assay results is determined by the performance of test kit manufacturer supplied reagents labeled as Acontrols@, used in accordance with the test kit instructions. Historically, FDA has not required additional assay-specific quality control procedures beyond those documented in the test kit package inserts. FDA now recommends implementation of quality assurance and quality control procedures that involve use of external control reagents beyond those provided by the test kit manufacturer in order to contribute to overall testing accuracy and therefore to blood safety. This guidance finalizes the draft guidance entitled AGuidance for Industry-Revised Recommendations for the Invalidation of Test Results When Using Licensed and 510(k) Cleared Bloodborne Pathogen Assays to Test Donors@ that was announced in the Federal Register of September 1, 1999 (64 FR 47847). This guidance also supercedes the January 3, 1994, guidance document on this topic (Ref. 1). These invalidation procedures should be part of blood establishments= comprehensive quality assurance programs (Ref. 2) and documented in blood establishments= Standard Operating Procedures (SOPs).

 

II. BACKGROUND

On May 28, 1992, September 25, 1992, September 26, 1996, and December 13, 1996, the Blood Products Advisory Committee (BPAC) discussed in open public meetings circumstances under which test results may be invalidated, including actions based on the use of additional quality control reagents in donor testing. These discussions included general aspects of quality assurance related to bloodborne pathogen testing, invalidation procedures based on test kit product insert requirements, and additional procedures to invalidate test results based on criteria developed by the user. FDA issued guidance on this subject on January 3, 1994 (Ref. 1). In an effort to further clarify FDA=s guidance and to help ensure compliance with the Clinical Laboratory Improvement Act of 1988 (CLIA), at the December 13, 1996, BPAC meeting, the FDA, the Centers for Disease Control and Prevention (CDC), and the Health Care Financing Administration (HCFA) discussed elements of CLIA. CDC described the differences between a calibrating reagent and a reagent serving as an assay control. Based on CLIA regulations, CDC stated that if a test kit supplied Acontrol@ reagent is used to calculate the assay cutoff, regardless of its container label title, the reagent serves the function of a calibrator and may not serve simultaneously as a control reagent. In such instances, an additional reagent should be included in the assay run to meet CLIA control reagent requirements.

 

For example, if a test kit includes two reagents labeled as negative control and positive control respectively, and the negative control reagent is used to calculate the assay cutoff, the negative control serves as a calibrator and may not also serve as an assay control reagent. Since CLIA regulations require use of a minimum of two controls (one negative and one positive) each day of testing, a negative control from an external source should be included in the run to meet CLIA requirements.

 

Such external controls may be developed in-house or procured, and the blood establishment must develop acceptance/rejection criteria for their use. For further information, blood establishments should refer to CLIA regulations under Title 42, Code of Federal Regulations (CFR), Part 493, or contact the State Survey Agency in the state where the laboratory is located or the accrediting organization with which the laboratory is affiliated.

 

It is important to note that the test kit package insert instructions must continue to be followed in order for runs to be considered valid for donor testing. 21CFR 606.65(e). Accordingly, externally supplied reagents should not be substituted for test kit supplied reagents to calculate the assay cutoff, but may be used additionally to satisfy CLIA requirements.

 

As a result of discussions held during BPAC meetings, and additional discussions among CDC, HCFA, and FDA, FDA has developed revised recommendations for blood establishments, integrating current CLIA provisions for invalidating test results based on CLIA designated control reagents.

 

Because blood safety relies significantly on the accuracy of testing, user-developed acceptance/rejection rules should be designed only with the objective of enhancing blood safety. Accordingly, rules developed by blood establishments to meet either CLIA requirements or other in-house developed requirements should only be used to invalidate nonreactive results. Reactive results obtained from a screening test or confirmatory/supplemental test run that satisfies the test kit== s criteria for acceptance should not be invalidated. Such reactive results remain as the initial test of record, and, in the case of a screening test, these specimens should be retested in duplicate as specified in the test kit manufacturer=s instructions. The duplicate retest of a reactive specimen (duplicate rescreen) is designed to reduce the probability of obtaining a false positive result for a truly negative, but non-specifically reactive specimen. A donor sample with an invalidated nonreactive result may be retested singly.

 

III. SPECIFIC RECOMMENDATIONS

These recommendations for procedures to invalidate test results, based on: 1) test kit package insert rejection criteria; and 2) CLIA provisions, are applicable to all blood establishments and should be included in their SOPs.

 

A. CLIA Requirements for Control Reagents

CLIA regulations require control reagents to be used according to 42 CFR 493. If a

bloodborne pathogen test kit uses any of its manufacturer supplied reagents to serve as a

calibrator function, i.e., either or both of the test kit controls (negative or positive) are used to calculate the assay cutoff, then CLIA regulations require that (an) additional Acontrol@ reagent(s) be included in each run (see table below). Such reagents may be procured or developed inhouse.

 

In any case, prior to placing the additional controls in routine use, each lot of such

reagents should have: 1) a known dating period, i.e., established stability (supplied by a control reagent manufacturer or established by the user on in-house developed control reagent); and 2) known performance parameters, i.e., specifications for acceptance. Prior to implementation, additional control reagents should be qualified, i.e., evaluated for suitability by the user applying generally accepted laboratory quality control procedures to establish acceptance/rejection criteria, to minimize possible incompatibilities that may exist with particular test kits before putting control reagents into use.

TEST KIT REAGENT(S) USED IN CALCULATION OF THE CUTOFF?

ADDITIONAL CONTROLREAGENT REQUIRED

 

 

B. Invalidation of Test Results Based on Test Kit Package Insert

 

 

All results, both reactive and non-reactive, must be considered invalid when it has been

determined that the assay, or part of the assay, has not been performed according to the test kit package insert instructions, (e.g., improper procedure, compromised reagent, faulty equipment) or that the test kit package insert defined run acceptance criteria have not been met, (e.g., controls out of specified range) [21 CFR 606.65(e)]. Whenever such situations occur, test results must be invalidated and the subsequent valid assay becomes the initial test of record.

 

The determination of nonadherence to the test kit package insert requirements may be a result of an investigation into an unexplained discrepancy, a routine review of the assay data, or concerns raised by a blood establishment's quality control procedures. For example, an investigation of procedures may be performed in response to unexpected external control reagent values or failure to meet the blood establishment's additional statistical acceptance criteria for rate of reactivity. Documentation of each incident of invalidation should include the basis for invalidation, the details of an investigation, including records of supervisory oversight, the outcome of the investigation, and, if indicated, any corrective action taken. All of these actions should be taken prior to repeat testing of donor samples.

 

C. Invalidation of Test Results Based on CLIA Regulations When the test kit package insert instructions have been met, but CLIA control requirements have not been met: 1) any reactive results must not be considered invalid and 2) nonreactive results should be invalidated. All specimens giving initially reactive results should be re-tested in duplicate, or as the package insert directs [21 CFR 606.65(e)]. If either of the repeated duplicates is reactive, the unit should be classified as repeatedly reactive and no further repeat testing should be performed. Specimens giving repeatedly reactive test results then should be further assayed and reported as recommended in FDA Recommendations to Blood Establishments (Ref. 3, 4, 5, 6, 7, 8, 9) regarding viral marker testing of donations. Invalidated non-reactive specimens should be re-tested singly and those results, if valid, should become the test of record.

 

Invalidation criteria pursuant to CLIA requirements should be based on site-user developed rules, e.g., population reactive prevalence data, or standard deviation based limits, which may be derived from generally accepted clinical laboratory quality control algorithms.

 

Documentation for all incidents of invalidation should include the basis for invalidation, the details of an investigation, including records of supervisory oversight, the outcome of the  investigation, and, if indicated, any corrective action taken. All of these actions should be taken prior to repeat testing of donor samples.

 

Reactive results should only be invalidated when an assay run either fails to meet test kit

package insert acceptance criteria or the assay was not performed in accordance with the test kit package insert. Accordingly, invalidation of reactive test results should not be based solely on either CLIA required quality control reagent performance or site supplemental quality control rules. Examples of circumstances when reactive results should not be invalidated include CLIA required control reagent values exceeding site established quality control acceptance limits, or reactive rates exceeding site defined upper limits.

IV. REFERENCES

1. FDA Recommendations for the Invalidation of Test Results When Using Licensed Viral Marker Assays to Screen Donors, 3 January 1994, Center for Biologics Evaluation and Research,  FDA.

2. Guideline for Quality Assurance in Blood Establishments, 11 July 1995, Center for Biologics Evaluation and Research, FDA.

3. Revised Recommendations for Testing Whole Blood, Blood Components, Source Plasma and  Source Leukocytes for Antibody to Hepatitis C Virus Encoded Antigen (Anti-HCV), 23 April 1992, Center for Biologics Evaluation and Research, FDA.

4. Revised Recommendations for Testing Whole Blood, Blood Components, Source Plasma and Source Leukocytes for Antibody to Hepatitis C Virus Encoded Antigen (Anti-HCV), 5 August

1993, Center for Biologics Evaluation and Research, FDA.

5. Additional Recommendations for Testing Whole Blood, Blood Components, Source Plasma and Source Leukocytes for Antibody to Hepatitis C Virus Encoded Antigen (Anti-HCV), 16 May 1996, Center for Biologics Evaluation and Research, FDA.

6. Recommendations for the Management of Donors and Units that are Initially Reactive for Hepatitis B Surface Antigen (HBsAg), 2 December 1987, Center for Biologics Evaluation and Research, FDA.

7. Revised Recommendations for the Prevention of Human Immunodeficiency Virus (HIV) Transmission by Blood and Blood Products, 23 April 1992, Center for Biologics  Evaluation and Research, FDA.

8. FDA Recommendations Concerning Testing for Antibody to Hepatitis B Core Antigen (Anti- HBc), 10 September 1992, Center for Biologics Evaluation and Research, FDA.

9. HTLV-I Antibody Testing, 29 September 1989, Center for Biologics Evaluation and  Research, FDA.


Your name:
Gravatar Preview
Your email:
(Optional) Email used only to show Gravatar.
Your website:
Title:
Comment:
Security Code
CAPTCHA image
Enter the code shown above in the box below
Add Comment   Cancel 
 
Recent Blog Comments
Minimize
Re: Earn Extra Money Donating Plasma If You Have Lupus Antibodies
My name is Michael and I'm looking in to helping my girlfriends financial issues with systematic Lupus... I have tried different methods to encourage her... Im very new to Lupus so I know that there is alot to learn... I could really care less if she works I'd much rather have her home so I can keep a watchful eye on her... But she is a hard headed survivor... She wants to get a job but is afraid of all that involves with telling an employer about Lupus and how it makes you feel and also the possibility of being rejected because of it... No one wants to be rejected period especially for something that they have no control of...

So I'm writing in hopes that someone will read this and understand... I'm looking for an easy way to help her feel better about herself and feel like she's putting into our relationship... Like I've mentioned before I could careless if she brings in money or not... I'm perfectly fine just knowing she loves me... I want to show her that I love her too...

----------------------
re:
I'm impressed. You're truly well informed and very intelligent. You wrote something that people could understand and made the subject intriguing for everyone. I'm saving this for future use.

Claire
www.imarksweb.net
Marks Web

----------------------
re:
love it! very interesting topics, I hope the incoming comments and suggestion are equally positive. Thanks for sharing information that is actually helpful.



wizard88
www.tulleeho.org

----------------------
Great article. I have this read and find very informative.
I completely agree with you. I have no point to raise in against of what you have said I think you explain the whole situation very well.

----------------------
re:
I’m impressed. Very informative and trustworthy blog does exactly what it sets out to do. I’ll bookmark your weblog for future use.

Joseph
www.joeydavila.com
Your online Library

----------------------
re:
This post I use my daily needed, by visiting this post I gain more information and by catching every details, it will become more faster. I'm really so happy that i found this post because by getting more information it is a good thing. This is the good reference and this is also usefull.

Lyka
www.pdfpal.org
Daily Guide

----------------------
re:
This post I use my daily needed, by visiting this post I gain more information and by catching every details, it will become more faster. I'm really so happy that i found this

post because by getting more information it is a good thing. This is the good reference and this is also usefull.

Lyka
www.pdfpal.org
Daily Guide

----------------------
re:
I'm impressed. You're truly well informed and very intelligent. You wrote something that people could understand and made the subject intriguing for everyone. I'm saving this for future use.

Claire
www.imarksweb.net
Marks Web

----------------------
Re: Hepatitis A Symptoms - Do I have Hepatitis A?
Sad to hear hope things get better

When you need Android information
Look Up
Android 4 You
www.android4you.com

----------------------
Re: Earn Extra Money Donating Plasma If You Have Lupus Antibodies
I would like to be considered for contributions toward the inclusion of thus study.

----------------------

If you have  an autoimmune, infectious, or other disease, researchers need your help.  Donate plasma and receive $200/donation for your time.  Visit www.idonateplasma.com for details.

  

Blog Search
Minimize
Sign-Up to Donate Plasma for $200+
Minimize
Paid Plasma Donor Program Announcements
Minimize
Toxoplasmosis Study

Diagnosed? Get $400/Hour for your time. http://www.idonateplasma.com

 read more ...

Celiac Disease Study

Diagnosed? Get $400/Hour for your time. http://www.idonateplasma.com

 read more ...

Hepatitis A Study

Diagnosed? Get $400/Hour of your time. http://www.idonateplasma.com

 read more ...

Autoimmune Disease? Lupus | RA | Other? Get Paid for Plasma Donation

Diagnosed? Get $400/Hour of your time. http://www.idonateplasma.com

 read more ...

Plasma Donation News Feed
Minimize
 

Site Map      |      Privacy Policy      |      Legal Statement      |      www.seracarecatalog.com      |      508.244.6400      |      info@seracare.com

.

  |